Melting Point Apparatus LDMP-B20

Melting Point Apparatus LDMP-B20 offers desktop design, visual melting point apparatus is of reliable performance. The Temperature ranges from Room temperature to 280˚C and offers 0.1˚C minimum digital display. Designed with LCD display. Measurement of the melting point of medicines, dye-stuffs and perfumes and other organic crystal substances to determine purity. Measurement method is in conformity with the standard of pharmacopoeia. It Measures three samples at the same time and calculate average value of initial and final melting points automatically.

Buy Now View Catalog
Measurement range Room temperature to 280˚C
Heating conducting medium Silicon oil
Minimum digital display 0.1˚C
Linear heating up rate 0.5, 1.0, 1.5, 3.0 ˚C/min
Accuracy ≤ 200˚C, ± 0.4˚C, 200˚C, ±0.7˚C
Repeatability (Standard Deviation) 0.3˚C (heating rate 1.0˚C/min)
Size of capillary Φ 1.4 mm (outside diameter) Φ 1.0 mm (inside diameter) Length: 120 mm
Number of capillary 3
Loading height of Sample 3 mm
Display LCD
Interface RS232
Dimension 390 × 390 × 520 mm
Gross Weight 13 kg
  • Designed with LCD display
    Temperature ranges from Room temperature to 280˚C
    Equipped with heating conducting medium silicon oil
    Offers 3 mm loading height of sample
    It Measures three samples at the same time
    Automatically calculate average value of initial and final melting points

It finds application across finding meting point of oil and fat in food and drug industry, biochemistry, medicines, mining, water management industry.

{"Measurement range":"Room temperature to 280\u02daC","Heating conducting medium":"Silicon oil","Minimum digital display":"0.1\u02daC","Linear heating up rate":"0.5, 1.0, 1.5, 3.0 \u02daC\/min","Accuracy":"\u2264 200\u02daC, \u00b1 0.4\u02daC, 200\u02daC, \u00b10.7\u02daC","Repeatability (Standard Deviation)":"0.3\u02daC (heating rate 1.0\u02daC\/min)","Size of capillary":"\u03a6 1.4 mm (outside diameter) \u03a6 1.0 mm (inside diameter) Length: 120 mm","Number of capillary":"3","Loading height of Sample":"3 mm","Display":"LCD","Interface":"RS232","Dimension":"390 \u00d7 390 \u00d7 520 mm","Gross Weight":"13 kg"}
Designed with LCD display
Temperature ranges from Room temperature to 280˚C
Equipped with heating conducting medium silicon oil
Offers 3 mm loading height of sample
It Measures three samples at the same time
Automatically calculate average value of initial and final melting points
It finds application across finding meting point of oil and fat in food and drug industry, biochemistry, medicines, mining, water management industry.


Touch Screen Melting Point Apparatus

Melting Point Apparatus LDMP-A12

Temperature Range RT to 320°C
Cooling Function NO
Sample Cell Capillary tube

Melting Point Apparatus LDMP-A13

Temperature Range RT to 400°C
Cooling Function Built-in Cooling Fan
Sample Cell Capillary tube