Flammability Tester LFLC- A11 Catalog

Download Catalog

Flammability Tester LFLC- A11

Overview

Flammability Tester LFLC- A11 is a horizontal and vertical flame chamber with PLC touch screen control panel. The built-in thermocouple calibration and temperature control system maintains high degree of automation. It has automatic ignition and temperature function that display temperature and time. Applicable to the single wire and cable, and soft-line horizontal, vertical combustion test. Adjustable burning time and gas flow that can be automatic / manual, vertical / horizontal switching. Designed according to the relevant provisions of UL1581-2006 1080 (VW -1 combustion test) and 1100 section (horizontal sample/FT2 combustion test) standards. It is convenient to observe and reliable.

Features :

  • PLC touch screen large display
    Control system includes flow meter, combustion reducing valve and u-shaped differential pressure meter
    Test chamber consists of test part and control part
    Main fixture: Test stand and burning torch
    SUS 304 stainless steel stents and vertical point for test support
    Stainless steel coated K type thermocouple
    PID temperature controller with readings up to 1000°C
    Spraying of high anti-corrosion coat on steel seal prevents air flow while testing
    Adjustable flame height and stable flame burner in accordance with ASDM 5025-05
    Enclosure tough glass front door for observation
    Low noise industrial exhaust fan works rapidly, prevents explosion of person to gas completely
  • Specifications :
    Control system PLC touch screen display
    Flame angle Adjustable
    Flame height 20 to 110 mm/min (adjustable)
    Gas supply Methane or prepared gas
    Gas flowmeter 1000 ml / min
    Gas press Adjustable
    Air inlet height 110 mm
    Burn time 0 to 999 s (adjustable)
    Nozzle diameter 9.5 mm
    Nozzle length 1.65 mm
    Thermal sensor 10.0 ± 0.05 g and thermocouple electrolytic copper piece
    Thermocouple type K type thermocouple (with 0.5 mm diameter stainless steel pipe coating)
    Thermocouple diameter 4 mm
    Cabinet capacity 4 cubic meter with glass watch door
    Interior lamp source LED lamp
    Test flame rising temperature 102°C to 103°C in 56 s
    After flame time Adjustable
    Sample vertical moving distance Same as test standard UL1581
    Sample horizontal moving distance Same as test standard UL1581
    Test precision 0.01 s
    External dimension (W × D × H) 2100 × 1360 × 2200 mm
    Power 500 W
    Power supply AC 220 V, 50 / 60 Hz
    Net weight 300 kg
    Gross weight 400 kg