HPLC system LHLC-B11 Catalog

Download Catalog

HPLC system LHLC-B11

Overview

This HPLC system delivers high quality performance and reliable results in order to meet the stringent requirements of the customer. To meet GMP/GLP requirements of the instrument, this HPLC system has hardware with VP function which records maintenance information and operations.

Features :

  • High pressure constant-current infusion pump
    Even mixing of the injections using mixers
    Supreme appearance to meet the customer’s visual requirements
    Simple, easy to use operations
    UV detector unit consists of dual beam monochromator, light battery, deuterium lamp
  • Specifications :
    High pressure pump High pressure constant-current infusion pump
    Infusion Way Micro volume series dual-piston
    Infusion pressure 0 - 9999 Psi
    Flow rate ≤ 0.5 %
    Flow rate range 0.001 - 9.999 mL / min (in 0.001 mL / min increments)
    Stability error for the flow rate ≤ 0.2 % RSD
    Gradient error ± 1 % (0 - 100 % water/acetone solvent)
    Pressure pulsation < 15 Psi (flow rate 1 mL / min, pressure 600 - 1600 Psi)
    Air tightness Pressure 5400 Psi, 10 min time, pressure drop < 400 Psi
    Time program function Yes
    Size (W × H × D) 260 × 130 × 420 mm
    Weight 11 kg
    Operation temperature 4 - 40 ºC
    UV detector
    • Wavelength range: 240
    • Wavelength error: ≤ 1 nm
    • Wavelength repeatability error: ≤ ± 0.1 nm
    • Dynamic noise: ≤ ± 0.75 × 10-5 AU (methyl alcohol 1 mL / min, 254 nm, 20 ºC)
    • Static noise: ≤ ± 0.5 × 10-5 A (empty cell, response time, 1 second, 20 ºC)
    • Dynamic baseline Drift: ≤ ± 1 × 10-4 AU / h (methyl alcohol 1 mL / min, 254 nm, 20 ºC)
    • Static baseline Drift: ≤ 0.5 × 10-4 (empty cell, response time, 1 second, 20 ºC)
    • Linear range: ≥ 104
    • Limit of detection: ≤ 1 × 10-9 g / ml (Naphthalene methyl alcohol solvent)
    • Quantitative repeatability: RSD 6 ≤ 0.1 %
    • Qualitative repeatability: RSD 6 ≤ 0.5 %
    • Spectrum band width: 6 nm
    • Flow cell volume: 8 µL
    • Optical path: 10 nm
    • Time program function: Yes
    • Size (W × H × D): 260 × 130 × 420 mm
    • Weight: 11 kg
    • Operation Temperature: 4 - 40 ºC
    Applications :

    This HPLC has applications in the scientific research studies, detection of residual amount in medicine, food, alcohol and in chemical analysis.