Atomic Absorption Spectrophotometer LAAS-A13

Flame Atomic Absorption Spectrophotometer LAAS-A13 is a microprocessor controlled instrument based on analysis of metal ions by flame and graphite furnace atomizer with their absorption at wavelength of 190 to 900 nm. Czerny-Turner monochromator and variance of bandwidth provides user flexibility. It is provided with deuterium hollow cathode lamp to exclude interference of molecular absorption and Titanium Burner to increase the analysis efficiency and precision of flame. Its double beam radiation sources enables good basic line stability. Graphic analysis system provides data on analysis of entire process e.g. measured value, temperature, procedures, time, etc. It has great applications in research, medicine, environmental science and other industries.

Buy Now View Catalog
Working spectral range 190 to 900 nm
Spectral Bandwith 0.1, 0.2, 0.4, 1.0 and 2.0 nm
Wavelength accuracy ± 0.15 nm
Wavelength repeatability ≤ 0.04 nm
Baseline stability ≤ 0.002 Abs / 30 min
Characteristic concentration of copper 0.02 μg / ml / 1 %
Detection limit of copper 0.004 μg / ml
Measurement Repeatability 0.005
Grating 1800 lines / mm
Inflamer All-metal titanium burner
Atomizer Efficient glass atomizer
Lamp stand 8 lamps automatic turret, automatic alignment
D2 Background calibration ability 1 A ≤ 30 times correction ability
Dimensions 700 × 550 × 450 mm
Net weight 75 Kg
Power supply 220 V ± 22 V AC
Acetylene air burner 100 mm
Ignition dynamic baseline drift ≤ 0.006 Abs / 30 min
Characteristic viscocity ≤ 0.025 μg / ml / 1 %
Standard deviation of the accuracy ≤ 0.5 % (Cu, absorbance > 0.8 A) (detection limit Cu ≤ 0.008 μg / ml)
Protection system Can inevitably cut off gas if pressure is low, power cut-off, flame out and alteration of the burner
  • Integrated flame/graphite furnace atomization system, changeable with flame emission burner
    Turret for 8 pre-aligned Hollow Cathode Lamps fitted with pre-heating, and individual pre-centred lamps
    PID and Dual curve mode light-controlled temperature control technique ensures fast heating and analytical sensitivity
    Fully automated wavelength scanning and peak searching
    Advance software for complete parameter settings
    Reliable fully automatic graphite furnace analysis
    Alarm feature to protect fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system
    Analysis of elements by flame absorption method and emission method
    Strong database system which possesses more than 500 data self-storage
    Results printed as parameter, data result and diagram
Accessories no. Name
1 PC workstation
2 Printer
3 Oil free Air compressor
4 Acetylene reducing valve
5 Cu - Hollow cathode lamp
6 Air filter
GRAPHITE FURNACE
Maximum temperature 3000 ℃
Temperature increasing speed ≥ 2000 ℃ / S
Characteristic quantity (Cd) ≤ 0.5 × 10 to 12g , (Cu) ≤ 0.5 × 10 to 11g
Accuracy (Cu) ≤3 % , (Cd) ≤ 3 %
Dimension 550 × 450 × 300 mm
Net weight 65 Kg
Safety system Protection for Overload current, reduced air pressure, low cooling water flow
Power supply 220 V ± 22 V AC, 7000 W

It is suitable to analyse trace amount of element from mixture, also in forensics and chemistry, biological science to detect toxic elements from blood and drug samples.

{"Working\u00a0spectral\u00a0range":"190 to 900 nm","Spectral Bandwith":"0.1, 0.2, 0.4, 1.0 and 2.0 nm","Wavelength accuracy":"\u00b1 0.15 nm","Wavelength repeatability":"\u2264 0.04 nm","Baseline\u00a0stability":"\u2264 0.002 Abs \/ 30 min","Characteristic\u00a0concentration\u00a0of\u00a0copper":"0.02 \u03bcg \/ ml \/ 1 %","Detection\u00a0limit\u00a0of\u00a0copper":"0.004 \u03bcg \/ ml","Measurement Repeatability":"0.005","Grating":"1800 lines \/ mm","Inflamer":"All-metal titanium burner","Atomizer":"Efficient glass atomizer","Lamp\u00a0stand":"8 lamps automatic turret, automatic alignment","D2\u00a0Background\u00a0calibration\u00a0ability":"1 A \u2264 30 times correction ability","Dimensions":"700 \u00d7 550 \u00d7 450 mm","Net weight":"75 Kg","Power supply":"220 V \u00b1 22 V AC","Acetylene air burner":"100 mm","Ignition dynamic baseline drift":"\u2264 0.006 Abs \/ 30 min","Characteristic viscocity":"\u2264 0.025 \u03bcg \/ ml \/ 1 %","Standard deviation of the accuracy":"\u2264 0.5 % (Cu, absorbance > 0.8 A) (detection limit Cu \u2264 0.008 \u03bcg \/ ml)","Protection system":"Can inevitably cut off gas if pressure is low, power cut-off, flame out and alteration of the burner"}
Integrated flame/graphite furnace atomization system, changeable with flame emission burner
Turret for 8 pre-aligned Hollow Cathode Lamps fitted with pre-heating, and individual pre-centred lamps
PID and Dual curve mode light-controlled temperature control technique ensures fast heating and analytical sensitivity
Fully automated wavelength scanning and peak searching
Advance software for complete parameter settings
Reliable fully automatic graphite furnace analysis
Alarm feature to protect fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system
Analysis of elements by flame absorption method and emission method
Strong database system which possesses more than 500 data self-storage
Results printed as parameter, data result and diagram
It is suitable to analyse trace amount of element from mixture, also in forensics and chemistry, biological science to detect toxic elements from blood and drug samples.


Related Products of Atomic Absorption Spectrophotometer

Labtron Atomic Absorption Spectrophotometer makes use of the modern in technology. It is acquired after thorough research into the desires of the diagnostic community. It has all the power, performance and speed necessitated in today’s modern laboratory.

Atomic Absorption Spectrophotometer LAAS-A22

Atomic Absorption Spectrophotometer LAAS-A22

Wavelength range : 180 – 900 nm

Spectral Bandwidth : 0.1 nm,0.2 nm,0.4 nm, 1.0 nm, 2.0 nm automatic setting

Wavelength accuracy : ≤ 0.15 nm

Atomic Absorption Spectrophotometer LAAS-A11

Atomic Absorption Spectrophotometer LAAS-A11

Wavelength range : 190 - 900 nm

Spectral Bandwidth : 0.2 nm, 0.4 nm, 0.7 nm, 1.4 nm, 2.4 nm, 5.0 nm

Wavelength accuracy : ≤ ± 0.5 nm

Atomic Absorption Spectrophotometer LAAS-A21
Atomic Absorption Spectrophotometer LAAS-A21

Wavelength range : 180 – 900 nm

Spectral Bandwidth : 0 nm, 0.2 nm, 0.4 nm, 1.0 nm, 2.0 nm (5 steps with automatic changeover)

Wavelength accuracy : ≤ 0.15 nm

Flame Atomic Absorption Spectrophotometer LAAS-A12
Flame Atomic Absorption Spectrophotometer LAAS-A12

Wavelength range : 190 nm ~ 900 nm

Spectral bandwidth : 0.1 nm, 0.2 nm, 1.0 nm, 2.0 nm

Wavelength accuracy : ± 0.25 nm

Atomic Absorption Spectrophotometer LAAS-A20

Atomic Absorption Spectrophotometer LAAS-A20

Wavelength range : 180 - 900 nm

Spectral Bandwidth : 0, 0.2, 0.4, 1.0, 2.0nm (Automatic setting)

Wavelength accuracy : ≤ 0.15 nm

Double Beam UV/Vis Spectrophotometer

Our range of Double Beam UV/VIS spectrophotometers offer long optical path which ensure high accuracy and stability. Its design measures the transmittance of the sample and solvent simultaneously. It is highly precise and performs powerful analysis.

Double Beam UV/Vis Spectrophotometer LUS-B30

Wavelength range 190 to 1100 nm
Spectral bandwidth 2 nm
Wavelength Accuracy ± 0.1nm @ 656.1 nm, ± 0.3 nm @ all

Double Beam UV/Vis Spectrophotometer LUS-B31

Wavelength range 190 to 1100 nm
Spectral bandwidth 1 nm
Wavelength Accuracy ± 0.1nm @ 656.1 nm, ± 0.3 nm @ all

NIR Spectrophotometer

NIR Spectrophotometer LNIR-A10
Wavelength Range 900 nm to 2500 nm
Wavelength Accuracy ≤ 0.2
Wavelength Reproducibility ≤ 0.05

Portable Spectrophotometer

Portable Spectrophotometer LSP-A31
Illumination 45/0 (45 ring shaped illumination,0° observation angle)
Repeatability Light splitting reflectivity:standard deviation within 0.08% Color values: ΔE*ab<=0.03 (After calibration, standard deviation of 30 measurements on test white board, 5 second intervals) Maximum:0.05
Sensor High sensitivity silicon photodiode

Portable Spectrophotometer LSP-A30

Illumination 45/0 (45 ring shaped illumination,0° observation angle)
Repeatability Light splitting reflectivity:standard deviation within 0.08% Color values: ΔE*ab<=0.03 (After calibration, standard deviation of 30 measurements on test white board, 5 second intervals) Maximum:0.05
Sensor High sensitivity silicon photodiode

Fluorescence Spectrophotometer

Fluorescence Spectrophotometer LFS-A10

Light source 150 W Xenon lamp
Excitation wavelength 200 nm to 900 nm
Emission wavelength 200 nm to 900 nm
Fluorescence Spectrophotometer LFS-A12
Light source 150 W Xenon lamp
Excitation wavelength 200 nm to 900 nm
Emission wavelength 200 nm to 900 nm

Nano Spectrophotometer

We produce a wide range of Nano spectrophotometer for qualitative and quantitative analysis of DNA/RNA in the sample. Our product is highly sensitive and requires only 0.5-2µl of sample for analysis.

Nano Spectrophotometer LNS-A21
Wavelength range 200 - 850 nm
Path length 1 mm, 2 mm, 5 mm, 10 mm
Wavelength accuracy ± 1 nm

Nano Spectrophotometer LNS-A10

Wavelength range 200 - 800 nm
Path length 0.2 mm (For high concentration measurement); 1.0 mm (For ordinary)
Wavelength accuracy ± 1 nm

Talk to us?