Wavelength range ( 3600 lines / mm ) |
180 to 500 nm |
Wavelength range ( 2400 lines / mm ) |
180 to 800 nm |
Temperature |
20 ~ 28 °C |
Elements per minute |
26 elements |
Humidity |
70% |
Focal length |
1000 mm |
Standard deviation |
RSD ≤ 1.0 % |
Correlation coefficient |
≥ 0.9998 |
Gas source |
Argon ( 99.99 % ) |
Plasma gas flow meter |
100 ~ 1000 L / h |
Auxillary gas flow meter |
10 ~ 100 L / h |
Carrier gas flow meter |
10 ~ 100 L / h |
Cooling water flow rate |
5 liters per min |
Cooling water temperature |
21 ~ 26 °C |
Water outlet diameter |
Ø 10 mm |
Line dispersion rate |
0.26 nm |
Resolving power |
≤ 0.008 nm |
Incident slit |
20 µm |
Co-axial sprayer diameter |
6 mm |
Coil diameter |
25 mm ( 3 turn ) |
Double room fog diameter |
34 mm |
Quartz tube torch diameter |
20 mm |
Mirror dimensions |
78 x 105 x 16 mm |
-
Scanning range variance ( 190 nm to 500 nm / 190 nm to 800 nm )
Rapid analysis – 26 elements per minute
Temperature range – 20 to 28 °C
Optical pathway – Czerny-turners
Low detection limit
Detection of multi-elements including non-metallic
Single and multi-element measurement method
Indication for wavelength error and repeatability
Range of measurement – common to trace
Used in clinical analysis for detection of metals in biological fluids , environmental analysis for trace element detection in water , soil , plants , pharmaceuticals for traces of catalysts and poison metals like Cd , Pb , industries for noble metal detection
{"Wavelength range ( 3600 lines \/ mm )":"180 to 500 nm","Wavelength range ( 2400 lines \/ mm )":"180 to 800 nm","Temperature":"20 ~ 28 \u00b0C","Elements per minute":"26 elements","Humidity":"70%","Focal length":"1000 mm","Standard deviation":"RSD\u00a0\u2264\u00a01.0 %","Correlation coefficient":"\u2265\u00a00.9998","Gas source":"Argon ( 99.99 % )","Plasma gas flow meter":"100 ~ 1000 L \/ h","Auxillary gas flow meter":"10 ~ 100 L \/ h","Carrier gas flow meter":"10 ~ 100 L \/ h","Cooling water flow rate":"5 liters per min","Cooling water temperature":"21 ~ 26 \u00b0C","Water outlet diameter":"\u00d8\u00a010 mm","Line dispersion rate":"0.26 nm","Resolving power":"\u2264 0.008 nm","Incident slit":"20 \u00b5m","Co-axial sprayer diameter":"6 mm","Coil diameter":"25 mm ( 3 turn )","Double room fog diameter":"34 mm","Quartz tube torch diameter":"20 mm","Mirror dimensions":"78 x 105 x 16 mm"}
Scanning range variance ( 190 nm to 500 nm / 190 nm to 800 nm )
Rapid analysis – 26 elements per minute
Temperature range – 20 to 28 °C
Optical pathway – Czerny-turners
Low detection limit
Detection of multi-elements including non-metallic
Single and multi-element measurement method
Indication for wavelength error and repeatability
Range of measurement – common to trace
Used in clinical analysis for detection of metals in biological fluids , environmental analysis for trace element detection in water , soil , plants , pharmaceuticals for traces of catalysts and poison metals like Cd , Pb , industries for noble metal detection
Element
|
Be
|
Mg
|
Ca
|
Sr
|
Ba
|
Sc
|
Y
|
Eu
|
Yb
|
Wavelength ( nm )
|
313.042
|
279.553
|
393.366
|
407.771
|
455.403
|
335.373
|
371.03
|
381.967
|
369.419
|
Detection limit
|
< 1.0
|
< 1.0
|
< 1.0
|
< 1.0
|
< 1.0
|
< 1.0
|
< 1.0
|
< 1.0
|
< 1.0
|
Element
|
Lu
|
La
|
Gd
|
Dy
|
Ho
|
Er
|
Tm
|
Zn
|
Cd
|
Wavelength ( nm )
|
261.542
|
333.749
|
342.247
|
353.17
|
345.6
|
337.271
|
313.126
|
213.856
|
226.502
|
Detection limit
|
< 1.0
|
< 3.0
|
< 3.0
|
< 3.0
|
< 3.0
|
< 3.0
|
< 3.0
|
< 3.0
|
< 3.0
|
Element
|
Fe
|
Ti
|
Cu
|
Mn
|
V
|
B
|
Ni
|
Cr
|
Co
|
Wavelength ( nm )
|
239.562
|
344.941
|
324.754
|
257.61
|
309.311
|
249.773
|
221.647
|
205.552
|
238.204
|
Detection limit
|
< 3.0
|
< 3.0
|
< 3.0
|
< 3.0
|
< 3.0
|
< 5.0
|
< 5.0
|
< 5.0
|
< 5.0
|
Element
|
Zr
|
Mo
|
Ag
|
Au
|
Si
|
Al
|
Ir
|
P
|
Hg
|
Wavelength ( nm )
|
343.823
|
202.03
|
328.068
|
242.795
|
251.611
|
396.152
|
224.268
|
213.618
|
194.227
|
Detection limit
|
< 5.0
|
< 5.0
|
< 5.0
|
< 5.0
|
< 5.0
|
< 10.0
|
< 10.0
|
< 10.0
|
< 10.0
|